Appendix 5: Operation of the Shimadzu spectrophotometers

Appendix 5: Operation of the Shimadzu spectrophotometers Precautions: - Clean up spills in the sample compartment immediately. - Keep spectrophotometer vents (on sides of instrument) free of any obstructions. - Use clean, dry hands on the keypads. - Do not place anything on top of the spectrophotometer. Spectrophotometer setup - Remove dust cover from instrument and place in a clean, out of the way area. - Remove any cuvettes from the sample and reference beams. - Turn the POWER to ON (left side of instrument). Allow 15 min. for warm-up. - After initialization, the MAIN MENU will be displayed. Fixed wavelength measurements - Press [1] from MAIN MENU to enter Photometric mode - Select ABS or %T by pressing the [F1] key - Press the [GOTO WL] key, enter the desired wavelength, and press [ENTER] - Insert a sample blank into both the front and back slots and press the [AUTO ZERO] key. Be sure to wait until the machine beeps – it takes a minute. - Insert an unknown sample into the front slot and press [START] to read the sample absorbance (%transmittance) on the digital display. Single Scan - Set up the spectrophotometer as described in 1 above. Then: - Enter the Spectrum mode from the MAIN MENU. - Select the MODE (ABS, %T) by pressing [1]. - Select the scan speed: - Press [4] - Choose a speed (1-5) - Press [ENTER] - Enter wavelength range: - Press [2] - Enter upper limit - Press [ENTER] - Enter lower limit - Press [ENTER] - Set the (ABS, %T) range: - Press [3] - Enter upper limit - Press [ENTER] - Enter lower limit - Press [ENTER] - Perform a background correction - Place the appropriate blank and reference samples in position. - Press the [F1] key to perform a baseline correction. This may take a few minutes depending upon settings. - Set the sample and reference cells in place. - Press the [START] key to start scan. Time Drive - Set up the spectrophotometer as described in 1 above. Then: - Enter the Time Scan mode from the MAIN MENU - Select the MODE (ABS, %T) by pressing [1]. - Enter the desired wavelength by pressing the [GOTO WL] key, entering the wavelength, and pressing [ENTER]. - Set the (ABS, %T) range: - Press [3] - Enter upper limit - Press [ENTER] - Enter lower limit - Press [ENTER] - Set the time scale (sec, min) by pressing [4] - Set the measurement time by pressing [2]. The measurement interval will automatically be set relative to the length of measurement chosen. - Press the [START] key to start measurement. Shutting down - Turn the spectrophotometer POWER to OFF. - Replace dust cover on spectrophotometer and make sure all electronics and wires are covered.
← Previous Chapter Next Chapter →